IEC 62368-1 Annex D.1 Impulse Voltage Generator Test Equipment

Product Details:

Place of Origin: China
Brand Name: KINGPO
Certification: Calibration Certificate
Model Number: KP-1095S

Payment & Shipping Terms:

Minimum Order Quantity: 1
Price: To be quoted
Packaging Details: safety carton pack or plywood box
Payment Terms: T/T
Category:

Description

IEC 62368-1 Annex D.1 Impulse Voltage Generator Test Equipment

Product Information:

The impulse voltage generator conforms to the standards as follows.
IEC 61180-1:2016 < High-voltage test techniques for low voltage equipment - Part 1: Definitions, test and procedure requirements, test equipment> clause 7.
IEC60335-1:2016 clause 14.
IEC60950-1:2013 clause1.5.7.2, 1.5.7.3,2.10.3.9,6.2.2.1,7.4.2,7.4.3, G5, Annex N.
IEC60065:2014 clause13.3.4, Annex J.5, Annex K.

IEC 62368-1 Annex D.1 Circuit 1

Technical Parameters:

1. Voltage integrated wave: 10/700us (when the output of the waveform generator is open).

2. Integrated wave output impedance:40W±10%

3. Integrated wave output voltage waveform peak value: 0.1~6kV

4. Output polarity: positive, negative

5. Impulse output mode: automatic.

6. Impulse times: 1~9999 times

7. Impulse interval time: 30~99 seconds (please note: the minimum interval time for 10/700ms wave is 30s)

8. Dimensions: L500×W470×H260 mm

9. Weight: about 30kg

10. Output waveform definition

11. Voltage wave: Front time:T1=1.67T=10ms±30%; Time to half value:T2=700ms±20%

Important Safety Precautions

The impulse voltage generator generates high-voltage pulses during operation. For safety, please read this manual and use this equipment correctly. Please pay attention to the following points during use:

(1) Do not use this equipment when your hands are wet or the relative humidity exceeds 75%.

(2) Because there is a high-voltage pulse applied to the terminal (impulse out), please confirm that the high-voltage power supply is disconnected when you want to change the wiring (HV.POW light is off) before proceeding.

(3) The operation must strictly follow the instruction manual.

(4) Although electric shocks are rarely encountered when operating this equipment, for safety, do not eat during operation to prevent accidental obstruction of eating caused by electric shocks.

(5) This equipment uses a vacuum contactor to generate voltage pulses. To ensure the life of the vacuum switch, do not turn on the high voltage when the equipment is not being tested, especially do not continuously trigger the vacuum contactor under high pressure.

(6) With high voltage inside, please do not disassemble or open the shell to work at will.

(7) It contains high voltage, please do not disassemble or open the shell to work.

Configuration List

Item Name Quantity Unit Remark
Main device Impulse Voltage Generator 1 SET
Accessories Power cord 1 PC
Ground wire 1 PC 2 m
EUT output connection wire 2 PCS Gun type single head

Black and red wires

Terminals block 1 PC 6P
Coaxial cable 1 PC For connecting to Oscilloscope
Document User’s manual 1 copy
Warranty card 1 copy
 

Certificate of Quality

 

1 Copy
impulse test generator
Impulse Voltage Generator

Reviews

There are no reviews yet.


Be the first to review “IEC 62368-1 Annex D.1 Impulse Voltage Generator Test Equipment”